700B Headstage Calibration

點擊次數:1114 發布時間:2015-1-28

700B Headstage Calibration

The steps below provide a quick check of the calibration of the MultiClamp 700B.  It is assumed that appropriate shielding (as described in “Test the Noise", above) will be used during these tests.

1.  Connect an oscilloscope to the front panel PRIMARY or SCOPE Output BNC.

2.  Synchronize the oscilloscope by connecting to the rear panel SYNC OUTPUT BNC. Press the “Reset to Program Defaults" button in the MultiClamp 700B Commander to standardize the MultiClamp 700B.


50 G Range

1.  Press the “Options" button, choose the Gains tab, and select the 50 G feedback resistor in the Voltage Clamp pane.  Return to the main MultiClamp 700B Commander window.

2.  Plug the PATCH connector of the PATCH-1U model cell into the CV-7 headstage.

3.  Set Seal Test amplitude to 100 mV and frequency to 50 Hz, then check the box to make it active.

4.  Press Auto Cp Fast to remove the bulk of electrode capacitance transient.

5.  The resulting waveform should be square, except for an initial overshoot (possibly twice the size of the steady-state response) that settles to the baseline in about 1 to 2 ms.  The rise time to the peak of the overshoot should be about 50 μs.  The steady-state amplitude following the transient should be ~500 mVp-p (±50 mV).


5 G Range

1.  Change the feedback resistor to 5 G in the Options / Gains menu.

2.  Press Auto Cp Fast.

3.  The step response should be ~50 mVp-p (±5 mV).


500 M Range

1.  Press the “Reset to Program Defaults" button.  (By default, the 500 M

range is selected.)

2.  Check Seal Test and set the amplitude to 25 mV.

3.  Plug the CELL connector of the PATCH-1U model cell into the CV-7 headstage.

4.  Press the Auto Whole Cell button.

5.  Press Auto Cp Fast button.

6.  The step response should be ~25 mVp-p. 


50 M Range

1.  Change the feedback resistor to 50 M.

2.  Increase Output Gain to 10.

3.  Press Auto Whole Cell and Auto Cp Fast.

4.  The step response should be ~25 mVp-p.  



[ 打印 ] [ 返回頂部 ] [ 關閉

| 商鋪首頁 | 公司檔案 | 產品展示 | 供應信息 | 公司動態 | 詢價留言 | 聯系我們 | 會員管理 |
化工儀器網 設計制作,未經允許翻錄必究.Copyright(C) http://www.iqjiance.com, All rights reserved.
以上信息由企業自行提供,信息內容的真實性、準確性和合法性由相關企業負責,化工儀器網對此不承擔任何保證責任。
溫馨提示:為規避購買風險,建議您在購買產品前務必確認供應商資質及產品質量。
二維碼

掃一掃訪問手機站
主站蜘蛛池模板: 仁寿县| 怀集县| 潢川县| 锡林郭勒盟| 澄迈县| 安仁县| 加查县| 青阳县| 兴国县| 上饶市| 新巴尔虎左旗| 南昌市| 南皮县| 乃东县| 河津市| 奉化市| 颍上县| 光泽县| 上饶县| 东阳市| 正阳县| 南溪县| 临高县| 济源市| 巴林左旗| 客服| 汤阴县| 南雄市| 威海市| 穆棱市| 孟津县| 阿拉尔市| 德惠市| 望谟县| 遂溪县| 临清市| 胶州市| 福安市| 万源市| 兴安盟| 万宁市|