進口digisense溫度控制器快速操作手冊與售后服務

時間:2017-9-4閱讀:1709

Installation
♦ Install control in safe operating area.
♦ Plug Heater/Cooler into the output connector, located on the back of the control.
♦ Connect input sensor into the proper connection input on control and place the thermocouple in the process that is being controlled. Install ferrite on lead wire of sensor. If further instruction is needed, see the user manual on installation of the ferrite.
♦ Plug supplied AC cord into the IEC connector, located on back of the control.
Operation Set up
♦ To enter the user-configurable setup press the MENU key from the system status screen.
♦ Use the MENU key to advance through each menu setting.
♦ Use the UP & DOWN key to make changes in each menu setting screen.
♦ Reference the flow chart (see pages 16 & 17 of manual) to have visual of the overall menu layout.
♦ When all settings have been made and saved the control will return to the System Status screen.
♦ Set the Set Point temperature by pressing the select button one time. A line will appear under the Set Point value.

♦ Use the UP and DOWN arrow keys to change the Set Point value.
♦ Press the SELECT key one time when finished and the line under the Set Point will disappear.
♦ Press the RUN/STOP key when ready to start the heating/cooling operation.
♦ Running the TUNE feature is recommended when running a new process variables.

Tune Operation
♦ Set the process temperature as instructed in the operation setup.
♦ Verify that the Auto Tune feature is enabled in the menu settings.
♦ From the System Status screen set the Set Point temperature.
♦ Press the TUNE button and the indicator showing Tune will have be illuminated on display.
♦ The Set Point value isn’t changeable after the Tune process has started. The value is locked until the Tune process is complete or aborted by the user.
♦ Stopping the Tune operation prior to it finishing will cause the PID settings to be returned to factory default values.
♦ PID values will be saved in non-volatile memory.

Control Sensor
Proper placement of the sensor can eliminate many problems in the operation of the system. The probe should be placed so that it can detect any temperature change with little thermal lag. In a process that requires fairly constant heat output, the probe should be located close to the heater. In processes where heat demand is variable, the probe should be located close to the work area. Experimenting with probe location will provide optimum results for the users specific application.

會員登錄

X

請輸入賬號

請輸入密碼

=

請輸驗證碼

收藏該商鋪

X
該信息已收藏!
標簽:
保存成功

(空格分隔,最多3個,單個標簽最多10個字符)

常用:

提示

X
您的留言已提交成功!我們將在第一時間回復您~

以上信息由企業自行提供,信息內容的真實性、準確性和合法性由相關企業負責,化工儀器網對此不承擔任何保證責任。

溫馨提示:為規避購買風險,建議您在購買產品前務必確認供應商資質及產品質量。

撥打電話
在線留言
主站蜘蛛池模板: 安化县| 延寿县| 嘉祥县| 百色市| 富裕县| 沈丘县| 张家口市| 泰顺县| 贺州市| 甘孜| 公主岭市| 阳朔县| 石门县| 华坪县| 全州县| 靖边县| 襄樊市| 定西市| 乡宁县| 堆龙德庆县| 福清市| 阿拉善左旗| 道真| 乌兰浩特市| 阿拉善右旗| 南投市| 通化县| 沐川县| 高雄县| 怀远县| 丰顺县| 乐东| 金湖县| 枣庄市| 五华县| 天台县| 泽普县| 齐河县| 扎鲁特旗| 潼关县| 革吉县|